Integrating cine EPID, dynamic delivery, and the off-axis Winston-Lutz test to enhance quality control in multiple brain metastasis stereotactic radiotherapy

Phys Med. 2024 Apr:120:103343. doi: 10.1016/j.ejmp.2024.103343. Epub 2024 Mar 28.

Abstract

Purpose: Stereotactic radiotherapy (SRT) has transformed cancer treatment, especially for brain metastases. Ensuring accurate SRT delivery is crucial, with the Winston-Lutz test being an important quality control tool. Off-axis Winston-Lutz (OAWL) tests are designed for accuracy assessment, but most are limited to fixed angles and hampered by local-field shifts caused by suboptimal Multi-Leaf Collimator (MLC) positioning. This study introduces a new OAWL approach for quality control in multi-brain-metastasis SRT. Utilizing cine Electronic Portal Imaging Device (EPID) images, it can be used with dynamic conformal arc (DCA) therapy. However, dynamic OAWL (DOAWL) is prone to more local-field shifts due to dynamic MLC movements. A two-step DOAWL is proposed: step 1 calculates local-field shifts using dynamic MLC movements in the beam-eye view data from the Treatment Planning System (TPS), while step 2 processes cine EPID images with an OAWL algorithm to isolate true deviations.

Methods: Validation involved an anthropomorphic head phantom with metallic ball-bearings, Varian TrueBeam STx accelerator delivering six coplanar/non-coplanar DCA beams, cine EPID, and ImageJ's OAWL analysis algorithm.

Results: Inherent local-field shifts ranged from 0.11 to 0.49 mm; corrected mean/max EPID-measured displacement was 0.34/1.03 mm. Few points exceeded 0.75/1.0-mm thresholds.

Conclusions: This two-step DOAWL test merges cine-EPID acquisitions, DCA, OAWL, and advanced analysis and offers effective quality control for multi-brain-metastasis SRT. Its routine implementation may also improve physicist knowledge of the treatment precision of their machines.

Keywords: Cine EPID; Multiple brain metastases; Winston Lutz.

MeSH terms

  • Brain Neoplasms*
  • Diagnostic Imaging
  • Humans
  • Phantoms, Imaging
  • Quality Control
  • Radiosurgery*
  • Radiotherapy Dosage
  • Radiotherapy Planning, Computer-Assisted
  • Radiotherapy, Conformal* / methods
  • Radiotherapy, Intensity-Modulated*