Purpose: To evaluate the performance of a prototype flexible transbronchial cryoprobe compared with that of percutaneous transthoracic cryoablation and to define cone-beam computed tomography (CT) imaging and pathology cryolesion features in an in vivo swine model.
Materials and methods: Transbronchial cryoablation was performed with a prototype flexible cryoprobe (3 central and 3 peripheral lung ablations in 3 swine) and compared with transthoracic cryoablation performed with a commercially available rigid cryoprobe (2 peripheral lung ablations in 1 swine). Procedural time and cryoablation success rates for endobronchial navigation and cryoneedle deployment were measured. Intraoperative cone-beam CT imaging features of cryolesions were characterized and correlated with gross pathology and hematoxylin and eosin-stained sections of the explanted cryolesions.
Results: The flexible cryoprobe was successfully navigated and delivered to each target through a steerable guiding sheath (6/6). At 4 minutes after ablation, 5 of 6 transbronchial and 2 of 2 transthoracic cryolesions were visible on cone-beam CT. The volumes on cone-beam CT images were 55.5 cm3 (SE ± 8.0) for central transbronchial ablations (n = 2), 72.5 cm3 (SE ± 8.1) for peripheral transbronchial ablations (n = 3), and 79.5 cm3 (SE ±11.6) for peripheral transthoracic ablations (n = 2). Pneumothorax developed in 1 animal after transbronchial ablation and during ablation in the transthoracic cryoablation. Images of cryoablation zones on cone-beam CT correlated well with the matched gross pathology and histopathology sections of the cryolesions.
Conclusions: Transbronchial cryoablation with a flexible cryoprobe, delivered through a steerable guiding sheath, is feasible. Transbronchial cryoablation zones are imageable with cone-beam CT, with gross pathology and histopathology similar to those of transthoracic cryoablation.
Published by Elsevier Inc.