Effective surface disinfection is crucial for preventing the spread of pathogens in hospitals. Standard UltraViolet-C (UV-C) lamps have been widely used for this purpose, but their disinfection efficiency under real-world conditions is not well understood. To fill this gap, the influence of the power of the ultraviolet radiator, source-sample distance, and exposure time on the performance of UV-C lamps against Escherichia coli and Staphylococcus epidermidis were experimentally determined in the laboratory and hospital. The obtained results showed that the UV irradiance and, thus, the UV-C disinfection efficiency decreased significantly at distances greater than 100 cm from the UV-C lamp. Moreover, increasing the total power of the radiators does not improve the performance of UV-C lamps under real conditions. The UV-C disinfection efficiency greater than 90% was achieved only under laboratory conditions at a close distance from the UV-C lamp, i.e., 10 cm. These findings provide novel insights into the limitations of UV-C lamps in real-world conditions and highlight the need for more effective disinfection strategies in hospitals.
Copyright: © 2024 Arkusz et al. This is an open access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.