With the increased provision of services by health authorities and community organizations allowing supervised inhalation of illicit substances comes concerns about the potential for secondhand exposure to the substances being used, whether in the adjacent community or to workers at the sites. In order to address community concerns surrounding secondhand illicit substance exposure and better protect harm reduction workers, a validated sampling and LC-MS/MS analysis method was developed for 6 illicit drugs: fentanyl, heroin, methamphetamine, cocaine, etizolam, and bromazolam. It was found that the filter used needed to be silanized to be made more inert and avoid loss of analyte due to degradation. Using the silanized filters, recoveries were good (>90%) and the collected samples were found to be stable at room temperature for 2 wk. The sampling volume validated was up to 960 L. The sensitivity and range of the method make it appropriate for short-term (15 min), full shift (8 h), or environmental sampling.
Keywords: air sampling; environmental sampling; fentanyl; illicit drugs; occupational exposure; occupational hygiene.
© The Author(s) 2024. Published by Oxford University Press on behalf of the British Occupational Hygiene Society.