Electro-mechanical co-stimulation of cells can be a useful cue for tissue engineering. However, reliable co-stimulation platforms still have limitations due to low durability of the components and difficulty in optimizing the stimulation parameters. Although various electro-mechanical co-simulation systems have been explored, integrating materials and components with high durability is still limited. To tackle this problem, we designed an electro-mechanical co-stimulation system that facilitates uniaxial cyclic stretching, electrical stimulation, and optical monitoring. This system utilizes a robust and autoclavable stretchable multielectrode array housed within a compact mini-incubator. To illustrate its effectiveness, we conducted experiments that highlighted how electro-mechanical co-stimulation using this system can enhance the maturation of cardiomyocytes derived from human induced pluripotent stem cells. The results showed great potential of our co-stimulation platform as an effective tool for tissue engineering.
Copyright © 2024 A Ri Kim et al.