Evaluation of custom-made Relay® stent-grafts for aortic arch landing zones 0 and I: experience from two high-volume aortic centres

Eur J Cardiothorac Surg. 2024 Jul 1;66(1):ezae241. doi: 10.1093/ejcts/ezae241.

Abstract

Objectives: To report experience with a Relay® stent-graft custom-made platform in treating different aortic arch pathology in 2 high-volume aortic centres.

Methods: A retrospective analysis of all patients treated between July 2016 and July 2023 with custom-made Relay® stent-graft (custom-made device). Underlying aortic arch pathology was an aneurysm, penetrating aortic ulcer, and dissection. Three custom-made device designs were used: proximal scallop, fenestrations, and inner branches. The endpoints were technical success, perioperative stroke, death, and reintervention rate.

Results: Thirty-five patients (89% males) with a mean age of 70 ± 11 years were treated.Indication for treatment was penetrating aortic ulcer in 14 patients (40%), aneurysm in 11 patients (31%) and aortic dissection in 10 patients (29%). The technical success rate was 100%. Twenty-eight patients (80%) had proximal sealing in zone 0, and 7 (20%) had proximal sealing in zone 1. Nine patients (25.6%) had proximal scallops, 9 (25.6%) had 1 big fenestration and 17 (48.8%) had a branched device; 1 with single branch, 15 with double branches and 1 with triple branches. Thirty patients (86%) had previous or simultaneous left subclavian artery revascularization. No patient died during 30 days. Two patients (5.7%) had stroke postoperatively; both recovered without disabling deficits. The mean follow-up was 35 ± 26 months. Six patients (17.1%) died during follow-up. One patient required reinforcement of the bridging stent in the left common carotid artery and one additional vascular plugging of the left subclavian artery. Three patients received distal extension.

Conclusions: The Relay® stent-graft custom-made platform showed a good performance in our study with a high technical success rate, low perioperative stroke and mortality, and low reintervention rates during the follow-up.

Keywords: Aortic arch; Custom-made platform; Design diversity; Relay®; stent-graft.

Publication types

  • Multicenter Study

MeSH terms

  • Aged
  • Aged, 80 and over
  • Aorta, Thoracic* / diagnostic imaging
  • Aorta, Thoracic* / surgery
  • Aortic Diseases / surgery
  • Blood Vessel Prosthesis Implantation* / instrumentation
  • Blood Vessel Prosthesis Implantation* / methods
  • Blood Vessel Prosthesis*
  • Endovascular Procedures / instrumentation
  • Endovascular Procedures / methods
  • Female
  • Humans
  • Male
  • Middle Aged
  • Prosthesis Design*
  • Retrospective Studies
  • Stents*
  • Treatment Outcome