Aim: Volar locking plates designed for far distal radius fracture fixation can have a significant hardware removal rate and risk of tendon rupture. Plate design has a role in the rate of complications. This study assessed the hardware removal and tendon rupture rate of the Acu-Loc 2 volar distal radius (VDR) plate often used in the treatment of far distal radial fractures.
Method: We searched our electronic healthcare records system for all patients who had undergone fixation with an Acu-Loc 2 VDR plate (Acumed, Hillsboro, OR, USA) at a tertiary center between January 2017 and December 2021. Patients were excluded if their follow-up time was less than one year or if they could not be contacted by telephone follow-up. Pre-operative radiographs were examined for fracture classification. Follow-up time was defined as the last contact in the clinic or by telephone.
Results: A total of 92 patients underwent an open reduction and internal fixation (ORIF) with an Acu-Loc 2 VDR plate. A total of 85 patients met the inclusion criteria for this study. Our cohort included 33 males (38.8%) and 52 females (61.2%). The mean age was 50 years. Twenty-seven fractures (31.0%) were extra-articular, and 60 fractures (69.0%) were intra-articular. The mean follow-up time for the patients was 593.3 days (range 369 to 1185 days). Four patients (4.7%) had their hardware removed. Three (3.5%) patients underwent removal due to tendon irritation and one patient (1.2%) due to a peri-prosthetic fracture around the plate. There were no tendon ruptures recorded.
Conclusion: The Acu-Loc 2 VDR plate had a low medium-term hardware removal rate and no tendon ruptures. These rates are lower than would be expected when compared with other far distal plate designs.
Keywords: distal end radius fracture; distal end radius plating; distal locking plate; volar locking plate; wrist trauma.
Copyright © 2024, Bharadwaj et al.