Extrinsic outflow graft obstruction of the HeartMate 3 LVAD: A state-of-the-art review

J Heart Lung Transplant. 2024 Nov;43(11):1873-1880. doi: 10.1016/j.healun.2024.07.005. Epub 2024 Jul 15.

Abstract

While notable improvements in survival, the incidence of hemocompatibility-related adverse events, hospitalizations, and cost have been demonstrated with the only commercially available durable left ventricular assist device, a category of pump malfunctions characterized by outflow graft obstruction has been noted with broader use and clinical follow-up of recipients of this technology. Of particular concern is the accumulation of acellular biodebris between the outflow graft and bend relief covering the outflow graft at its origin with the pump (which we term extrinsic outflow graft obstruction at the bend relief). This process tends to be insidious, occurs late in the postoperative course, can be challenging to diagnose, and can result in significant morbidity and mortality. Herein, we provide a review of this complication and outline diagnostic, treatment, and preventive strategies.

Keywords: HeartMate 3; LVAD; low-flow alarms; outflow graft obstruction; pump malfunction.

Publication types

  • Review

MeSH terms

  • Heart Failure / surgery
  • Heart-Assist Devices* / adverse effects
  • Humans
  • Prosthesis Failure