Objective: To investigate the effect of Polyetheretherketone (PEEK) rod semi-rigid pedicle screw fixation system in lumbar spine non-fusion surgery.
Methods: A total of 74 patients with tow-level lumbar degenerative diseases who underwent surgery from March 2017 to December 2019 were divided into PEEK rod group and titanium rod group. In the PEEK rod group, there were 34 patients, including 13 males and 21 females, aged from 51 to 79 years old with an average of (62.4±6.8) years old;There were 1 patient of L1-L3 segments, 7 patients of L2-L4 segments, 20 patients of L3-L5 segments and 6 patients of L4-S1 segments. In the titanium rod group, there were 40 patients, including 17 males and 23 females, aged from 52 to 81 years old with an average of (65.2±7.3) years old;There were 3 patient of L1-L3 segments, 11 patients of L2-L4 segments, 19 patients of L3-L5 segments and 7 patients of L4-S1 segments. The general conditions of operation, such as operation time, intraoperative blood loss, postoperative drainage was recorded. The visual analogue scale (VAS) for low back pain and Oswestry disability index (ODI) were compared in preoperatively and postoperatively(3 months, 12 months and last follow-up) between two groups. The change of range of motion (ROM) was observed by flexion and extension x-ray of lumbar.
Results: All patients successfully completed the operation. The follow-up time ranged from 22 to 34 months with an average of(26.8±5.6) months. The operative time (142.2±44.7) min and intraoperative blood loss(166.5±67.4)ml in PEEK group were lower than those in titanium group [(160.7±57.3) min、(212.8±85.4) ml](P<0.05). There was no significant differences in postoperative drainage between the two groups (P>0.05). At the final follow-up visit, in PEEK group and titanium group VAS of low back pain[(0.8±0.4) points vs (1.0±0.5) points], VAS for leg pain [ (0.7±0.4) points vs (0.8±0.5) points] and ODI [(9.8±1.6)% vs (12.1±1.5)%] were compared with preoperative [ (5.8±1.1) points vs (6.0±1.1)points], [ (7.2±1.7) points vs (7.0±1.6) points], [(68.5±8.9)% vs(66.3±8.2)%] were significantly different(P<0.05). There was no significant difference in VAS scores between the two groups at each postoperative time point (P>0.05). At 3 months after surgery, there was no difference in ODI between the two groups (P>0.05). There were significant differences in ODI between PEEK group and titanium rod group at 12 months [(15.5±2.1)% vs (18.4±2.4)%] and at the last follow-up [(9.8±1.6)% vs (12.1±1.5)%] (P<0.05). The total range of motion (ROM) of lumbar decreased in both groups after surgery. At 12 months after surgery and the last follow-up, the PEEK group compared with the titanium rod group, the total range of motion of lumbar was statistically significant (P<0.05). The range of motion (ROM) of the fixed segments decreased in both groups after surgery. The ROM of the fixed segments in PEEK group decreased from (9.5±4.6)° to (4.1±1.9)° at the last follow-up (P<0.05), which in the titanium rod group was decreased from (9.8±4.3)°to (0.9±0.5)° at the last follow-up (P<0.05). The range of motion (ROM) of upper adjacent segment increased in both groups, there was statistical significance in the ROM of upper adjacent segment between the two groups at 12 months after surgery and the last follow-up, (P<0.05). There was no screw loosening and broken rods in both groups during the follow-up period.
Conclusion: The PEEK rod semi-rigid pedicle screw internal fixation system used in lumbar non-fusion surgery can retain part of the mobility of the fixed segment, showing comparable short-term clinical efficacy to titanium rod fusion. PEEK rod semi-rigid pedicle screw internal fixation system is a feasible choice for the treatment of lumbar spine degenerative diseases, and its long-term efficacy needs further follow-up observation.
Keywords: Clinical efficacy; Lumbar degenerative disease; Non-fusion; Polyetheretherketone rods.