Does the Addition of Low-Dose Antibiotics Compromise the Mechanical Properties of Polymethylmethacrylate (PMMA)?

Polymers (Basel). 2024 Aug 22;16(16):2378. doi: 10.3390/polym16162378.

Abstract

The increasing numbers of total joint replacements and related implant-associated infections demand solutions, which can provide a high-dose local delivery of antibiotics. Antibiotic-loaded bone cement (ALBC) is an accepted treatment method for infected joint arthroplasties. The mechanical properties of low-dose gentamicin-loaded bone cement (BC) in medium- and high-viscosity versions were compared to unloaded BC using a vacuum mixing system. As an additional control group, manual mixed unloaded BC was used. In a uniaxial compression test, ultimate compressive strength, compressive yield strength, and compression modulus of elasticity, as well as ultimate and yield strain, were determined according to ISO 5833-2022 guidelines. All groups exceeded the minimum compressive strength (70 MPa) specified in the ISO 5833 guidelines. Both ALBC groups showed a similar ultimate compressive and yield strength to the unloaded BC. The results showed that vacuum mixing increased the compression strength of BC. ALBC showed similar compressive strength to their non-antibiotic counterparts when vacuum mixing was performed. Added low-dose gentamicin acted as a plasticizer on bone cement. From a biomechanical point of view, the usage of gentamicin-based ALBC formulations is viable.

Keywords: antibiotic-loaded bone cement; compression testing; gentamicin; local delivery of antibiotics; mechanical properties; periprosthetic joint infection.

Grants and funding

This research received no funding. Bone cement as well as mixing systems were provided by Heraeus Medical GmbH. Publication costs were covered by Heraeus Medical GmbH.