Dosimetric comparison of hippocampal-sparing technologies in patients with low-grade glioma

Neurooncol Adv. 2024 Aug 6;6(1):vdae131. doi: 10.1093/noajnl/vdae131. eCollection 2024 Jan-Dec.

Abstract

Background: Radiotherapy (RT) plays an integral role in the management of low-grade gliomas (LGG). Late toxicity from RT can cause progressive neurocognitive dysfunction. Radiation-induced damage to the hippocampus (HCP) plays a considerable role in memory decline. Advancements in photon planning software have resulted in the development of multi-criteria optimization (MCO) and HyperArc technologies which may improve HCP sparing while maintaining planning target volume (PTV) target coverage.

Methods: Three planning methods for hippocampal sparing (HS) were compared, volumetric modulated arc therapy (VMAT) without HS (VMAT_noHS), VMAT with HS (VMAT_HS), MCO with HS (MCO_HS), and HyperArc with HS (HyperArc_HS).

Results: Twenty-five patients were identified. The contralateral HCP was spared in 16 patients and bilateral HCP in 9 patients with superiorly located tumors. All 3 HS planning techniques showed significant reductions in dose to the spared HCP in contralateral cases but only VMAT_HS and MCO_HS achieved this in bilateral cases (P < .008). Only MCO_HS was superior to VMAT_HS in lowering the dose to both contralateral HCP and bilateral HCP in all measured metrics (P < .008). PTV and OAR (organ at risk) dose constraints were achieved for all plans.

Conclusions: This retrospective dosimetric study demonstrated the feasibility of HS for low-grade glioma. All 3 HS planning techniques achieved significant dose reductions to the spared contralateral hippocampus, but only MCO_HS and VMAT_HS achieved this in bilateral cases. MCO was superior to other planning techniques for sparing both bilateral and contralateral hippocampi.

Keywords: hippocampal sparing; hyperarc; low-grade glioma; multi-criteria optimizer; photon planning.