Label-free monitoring of embolotherapy via catheter electrochemical impedance spectroscopy

Sci Rep. 2024 Sep 12;14(1):21310. doi: 10.1038/s41598-024-71835-z.

Abstract

Catheter-based embolization has become a widely adopted minimally-invasive treatment for a broad range of applications. However, assessment of embolization endpoints requires x-ray fluoroscopic monitoring, exposing patients and physicians performing embolization procedures to harmful ionizing radiation. Moreover, x-ray fluoroscopy assessment of embolization endpoints is low sensitivity, subjective, and may not reflect the actual physiology of blood flow reduction, thus providing little oversight of the embolization procedure. Inspired by the observation that the dielectric properties of blood differ from those of fluids injected during the embolization procedure, a customized angiographic catheter was created with embedded electrodes for catheter-based electrochemical impedance spectroscopy as a way to monitor embolization. Real-time electrochemical impedance spectroscopy was performed in a phantom and compared to visual and videographic monitoring. Electrochemical impedance spectroscopy was able to sense endpoints of embolization, including stasis, reflux, and persistent flow. This new technique offers a label-free method of sensing embolization progress with potentially higher sensitivity and reproducibility compared to x-ray, as well as offer substantial reduction in x-ray exposure to patients and physicians.

Keywords: Electrochemical impedance; Embolization; In vivo sensing; Radiation exposure.

MeSH terms

  • Catheters*
  • Dielectric Spectroscopy* / methods
  • Electrodes
  • Embolization, Therapeutic* / methods
  • Humans
  • Phantoms, Imaging