Purpose: Optimizing the preparation of a 10 ml syringe for manual injection of contrast media can help operators obtain easier and faster injections. This study aims to compare the flow rates of different contrast media injection methods.
Methods: Different contrast media solutions were compared: 100% contrast (10 ml contrast), mixed contrast solution (8:2 contrast/saline), and layered contrast below saline ("Parfait") in different volumes. Contrast media were injected at room temperature (20°C) and after heating (37°C). Four operators injected 10 ml syringes filled with different mediums through 5-French angiographic catheters. The average flow rate was used to compare different contrast injection mediums. The Kruskal-Wallis test with post-hoc pairwise comparisons using Bonferroni correction or Mann-Whitney U-tests were employed depending on the type of comparison.
Results: Compared to the 100% contrast solution, every Parfait media and mixed contrast solution demonstrated significantly higher flow rates (p < 0.001). The 5 ml saline Parfait had the highest flow rate among the Parfait solutions. The 5 ml saline Parfait and the mixed solution had comparable flow rates (p = 0.237). Higher flow rates were observed upon heating both 100% contrast (p < 0.001) and mixed contrast solutions (p < 0.001) in comparison to their flow rates at room temperature.
Conclusion: This study demonstrates the capability of the Parfait and mixed contrast injections to achieve higher flow rates than the 100% contrast solutions. Heating the contrast media to 37°C also proves to be a viable strategy for further enhancing the flow rates for 100% and mixed contrast solutions.
Keywords: Contrast media; angiography; ergonomics.