Background: The efficacy and safety of the ultrathin BioMime sirolimus-eluting coronary stent (SES) system in treating single or multiple de novo native coronary lesions, in-stent restenosis, and bifurcation lesions have been evidenced at 1 year.
Aims: We sought to investigate the long-term safety and efficacy of the BioMime SES in a real-world population with obstructive coronary artery disease (CAD).
Methods: The prospective, single-arm, multicentre meriT-2 trial enrolled 250 patients from 11 sites across India. The safety endpoint was the cumulative frequency of major adverse cardiovascular events (MACE) at 5 years, defined as a composite of cardiac death, myocardial infarction (MI), emergent coronary artery bypass grafting or clinically indicated target lesion revascularisation (CI-TLR). Stent thrombosis (ST) was evaluated according to the Academic Research Consortium definitions.
Results: A total of 214 (85.6%) subjects completed the 5-year follow-up. The mean age of patients was 57.44±10.75 years, and 82.71% were males. A total of 308 lesions were treated with BioMime SES. Most of the lesions were localised in the left anterior descending artery (45.46%) and were type B2 lesions (44.81%). The cumulative MACE rate at 5 years was 8.9% (n=19), including 0.9% cardiac deaths, 1.9% MI and 6.1% CI-TLR. The rate of ST was only 0.5%. The Kaplan-Meier survivor analysis revealed actuarial survivorship of 95.6% for the intention-to-treat population (n=250) over 5 years.
Conclusions: The long-term clinical outcomes of the meriT-2 trial established the safety and efficacy of the ultrathin-strut biodegradable-polymer-based BioMime SES with satisfactory clinical outcomes at 5 years.