The LET enhancement of energy-specific collimation in pencil beam scanning proton therapy

J Appl Clin Med Phys. 2025 Jan;26(1):e14477. doi: 10.1002/acm2.14477. Epub 2024 Dec 7.

Abstract

Purpose: To computationally characterize the LET distribution during dynamic collimation in PBS and quantify its impact on the resultant dose distribution.

Methods: Monte Carlo simulations using Geant4 were used to model the production of low-energy proton scatter produced in the collimating components of a novel PBS collimator. Custom spectral tallies were created to quantify the energy, track- and dose-averaged LET resulting from individual beamlet and composite fields simulated from a model of the IBA dedicated nozzle system. The composite dose distributions were optimized to achieve a uniform physical dose coverage of a cubical and pyramidal target, and the resulting dose-average LET distributions were calculated for uncollimated and collimated PBS deliveries and used to generate RBE-weighted dose distributions.

Results: For collimated beamlets, the scattered proton energy fluence is strongly dependent on collimator position relative to the central axis of the beamlet. When delivering a uniform profile, the distribution of dose-average LET was nearly identical within the target and increased between 1 and 2 keV / μ m $2 \,{\rm keV}/\mathrm{\umu }\mathrm{m}$ within 10 mm surrounding the target. Dynamic collimation resulted in larger dose-average LET changes: increasing the dose-average LET between 1 and 3 keV / μ m $3 \,{\rm keV}/\mathrm{\umu }\mathrm{m}$ within 10 mm of a pyramidal target while reducing the dose-average LET outside this margin by as much as 10 keV / μ m $10 \,{\rm keV}/\mathrm{\umu }\mathrm{m}$ . Biological dose distributions are improved with energy-specific collimation in reducing the lateral penumbra.

Conclusion: The presence of energy-specific collimation in PBS can lead to dose-average LET changes relative to an uncollimated delivery. In some clinical situations, the placement and application of energy-specific collimation may require additional planning considerations based on its reduction to the lateral penumbra and increase in high-dose conformity. Future applications may embody these unique dosimetric characteristics to redirect high-LET portions of a collimated proton beamlet from healthy tissues while enhancing the dose-average LET distribution within target.

Keywords: LET; PBS; RBE; collimation; protons.