A novel focal lattice-tip catheter toggling between pulsed field energy and radiofrequency for atrial arrhythmia ablation: Results from a real-world, multicenter registry

Heart Rhythm. 2024 Dec 16:S1547-5271(24)03694-4. doi: 10.1016/j.hrthm.2024.12.021. Online ahead of print.

Abstract

Background: A novel focal lattice-tip catheter allowing the delivery of either pulsed field (PF) or radiofrequency (RF) energy has recently received regulatory approval. The technology features a proprietary 3-dimensional electroanatomic mapping system.

Objective: We describe the first real-world and multicenter experience.

Methods: Consecutive AF patients undergoing first-time or redo atrial tachyarrhythmia ablation with the Affera system were prospectively enrolled at 3 different centers. PF was the only energy source allowed when ablating the posterior left atrium; anterior applications were performed with either RF (PF/RF strategy) or PF (PF/PF strategy) on the basis of the operator's preference. The primary efficacy end point included acute electrical isolation of pulmonary veins and posterior wall or bidirectional block in case of linear lesions.

Results: The study included 130 patients (mean age, 67 ± 10 years; 63.8% [n = 83] male; 61.5% [n = 80] nonparoxysmal AF; 55.4% first-time AF ablation). First-time pulmonary vein isolation was performed in 72 patients: RF/PF in 13 (18.1%) patients and PF/PF in the remaining 59 (81.9%); first-pass isolation for pulmonary vein and posterior wall was achieved in 100% of cases. A total of 289 ablation lines were performed (roof line, 91 patients; inferior line, 83 patients; anterior mitral line, 32 patients; posterior mitral line, 45 patients; cavotricuspid isthmus line, 38 patients). First-pass isolation and primary efficacy end point were 96.2% (roof line, 100%; inferior line, 100%; anterior mitral line, 96.9%; posterior mitral line, 84.4%; cavotricuspid isthmus, 92.1%) and 100%, respectively. We had 2 (1.5%) major complications: 1 ST-segment elevation at the inferolateral leads requiring intracoronary administration of nitrate and 1 complete atrioventricular block.

Conclusion: Catheter ablation with a novel 9-mm lattice-tip catheter confirmed high efficacy and safety in a real-world scenario.

Keywords: Atrial fibrillation; Catheter ablation; Cavotricuspid isthmus; Lattice tip; Mitral isthmus; Pulmonary vein isolation; Pulsed field ablation; Roof line; Temperature controlled.