Comparison of single- and multi-isocenter planning with Dynamic WaveArc for multiple brain metastases

J Radiat Res. 2024 Dec 27:rrae098. doi: 10.1093/jrr/rrae098. Online ahead of print.

Abstract

Dynamic WaveArc (DWA) is a technique used for continuous, non-coplanar volumetric-modulated arc therapy on the Vero4DRT platform. This study aimed to evaluate the application of single-isocenter DWA (SI-DWA) for treating multiple brain metastases by comparing dose distribution and irradiation time with multi-isocenter DWA (MI-DWA) through retrospective treatment planning. Treatment plans were developed for SI-DWA and MI-DWA in 14 cases with 3-5 brain metastases. Parameters assessed included target dose indices, such as conformity index (CI) of the planning target volume (PTV), volumes of normal brain excluding gross tumor volumes (GTVs) receiving a single dose equivalent of 14 Gy (V14), V30%, V20%, V10%, volumes of normal brain, including GTVs receiving a single dose equivalent of 12 Gy (V12), D2% for other organs at risk, and beam-on time. SI-DWA showed inferior CI, V14, and V12 values for lesions with PTV volumes <1 cc, whereas it performed equivalently to MI-DWA for lesions with PTV volumes ≥1 cc. SI-DWA resulted in higher volumes of normal brain receiving low doses compared to MI-DWA. SI-DWA exhibited significantly shorter beam-on times than MI-DWA. In conclusion, SI-DWA is an effective method for treating multiple brain metastases with PTV volumes ≥1 cc, offering an index of radiation-induced brain necrosis comparable with MI-DWA while allowing for shorter irradiation times.

Keywords: Dynamic WaveArc; brain metastases; multi-isocenter; single-isocenter; stereotactic irradiation.