This study presents a comprehensive evaluation of the application of online multi-internal standard calibration (M.ISC) in determining iodine concentrations through inductively coupled plasma mass spectrometry (ICP-MS). Notably, M.ISC streamlines the calibration process by requiring only a single standard solution, thereby enhancing sample throughput and minimizing liquid waste. In addition, unlike conventional internal standard (IS) methods, M.ISC omits the need for time-consuming species identification by utilizing multiple IS species simultaneously to minimize signal biases. The effectiveness of M.ISC was validated through the analysis of six standard reference samples, with the results of LOD and LOQ also being calculated by the error propagation approach. The traditional chemical analytical methods (TCAM), external standard calibration (EC) and single IS methods were also evaluated as comparative purpose. Nonetheless, M.ISC emerges as a straightforward matrix-correction strategy, offering a simple and efficient alternative to traditional calibration methods for iodine detection by ICP-MS.
Keywords: ICP‐MS; arsenic–cerium catalytic redox method; iodine; matrix effects; multi‐internal standard calibration.