An MR imaging technique has been developed producing head and body images of diagnostic quality in only a few seconds acquisition time. The Fourier type imaging technique uses excitation with relatively small excitation angels, echoes produced by gradient inversion, and extremely fast profile repetition. A typical result at 0.5 T is an artifact-free head image of 128 x 128 resolution, 10 mm slice thickness in an acquisition time of 2 seconds.