Comparison of gas chromatographic--electron-capture detection and high-performance liquid chromatography for the determination of butofilolol in biological fluids

J Chromatogr. 1983 Dec 9;278(2):301-9. doi: 10.1016/s0378-4347(00)84789-1.

Abstract

Two selective and sensitive methods for the quantitative analysis of butofilol in human plasma and urine are discussed. The first method is a gas chromatographic assay with electron-capture detection using extraction with toluene, several clean-up procedures and derivatization. The second method is based on high-performance liquid chromatography and a single extraction with dichloromethane. The two assay methods were applied to the determination of the same human plasma samples after administration of a single, oral 200-mg dose of butofilolol. A good correlation between the results (inter-laboratory comparison) is obtained, validating both techniques.

Publication types

  • Comparative Study

MeSH terms

  • Chromatography, Gas / methods
  • Chromatography, High Pressure Liquid / methods
  • Humans
  • Propanolamines / analysis*
  • Propanolamines / blood

Substances

  • Propanolamines
  • butofilolol