Effects of analog and digital filtering on auditory middle latency responses in adults and young children

Ann Otol Rhinol Laryngol. 1984 May-Jun;93(3 Pt 1):267-70. doi: 10.1177/000348948409300316.

Abstract

Effects of analog high pass (HP) filtering were compared with those of zero phase-shift digital filtering on the auditory middle latency responses (MLR) from nine adults and 16 young children with normal hearing. Analog HP filtering exerted several prominent effects on the MLR waveforms in both adults and young children, such as suppression of Po (ABR), enhancement of Nb, enhancement or emergence of Pb, and latency decrements for Pa and the later components. Analog HP filtering at 20 Hz produced more pronounced waveform distortions in the responses from young children than from adults. Much greater latency decrements for Pa and Nb were observed for young children than for adults in the analog HP-filtered responses at 20 Hz. A large positive peak (Pb) emerged at about 65 ms after the stimulus onset. From these results, the use of digital HP filtering at 20 Hz is strongly recommended for obtaining unbiased and stable MLR in young children.

Publication types

  • Comparative Study

MeSH terms

  • Acoustics*
  • Adult
  • Child
  • Child, Preschool
  • Evoked Potentials, Auditory*
  • Humans
  • Reaction Time