A rapid radioimmunoassay method for serum luteinizing hormone utilizing polyethylene glycol and a double-antibody method of separation

Fertil Steril. 1981 Jan;35(1):36-9. doi: 10.1016/s0015-0282(16)45255-6.

Abstract

A method is described utilizing polyethylene glycol (PEG) combined with a double antibody for separation for rapid radioimmunoassay (RIA) of human luteinizing hormone(hLH). The total assay time is 3 hours with a 1-hour incubation period. Reliable comparisons with the 3-day assay have been shown. The average between-assay coefficient of variation was 11%, whereas the average within-assay coefficient of variation was 6%. The sensitivity of the assay was 7.5 mIU/ml. It is suggested that this reliable and rapid RIA for hLH will prove to be a valuable adjunct in the treatment of patients when timing of ovulation is imperative, such as for artificial insemination and harvesting of maturing human oocytes.