Improved detection of anti-HCV in post-transfusion hepatitis by a third-generation ELISA

Vox Sang. 1995;68(1):15-8. doi: 10.1111/j.1423-0410.1995.tb02538.x.

Abstract

The sensitivity of ORTHO HCV 3.0 ELISA Test System (ELISA 3) for the detection of anti-HCV was compared with the second-generation ELISA, OR-THO HCV 2.0 ELISA Test System (ELISA 2). ELISA 3 differs from ELISA 2 in that it incorporates the HCV recombinant antigen NS5, in addition to recombinant antigens derived from the NS3, NS4 and core regions of the HCV genome. Specimens tested consisted of serial bleeds obtained from 21 individuals undergoing seroconversion following acquisition of post-transfusion HCV infection. ELISA 3 demonstrated significantly greater sensitivity than ELISA 2, detecting seroconversion earlier in 24% (5/21) of cases. Although one of these cases appeared to represent early seroconversion to NS5, most of the improved sensitivity of ELISA 3 appeared to derive from increased detectability of anti-c33c.

MeSH terms

  • Enzyme-Linked Immunosorbent Assay*
  • Evaluation Studies as Topic
  • Hepacivirus / immunology
  • Hepacivirus / isolation & purification*
  • Hepatitis Antibodies / blood*
  • Hepatitis C / etiology
  • Hepatitis C / immunology*
  • Hepatitis C Antibodies
  • Humans
  • Sensitivity and Specificity
  • Transfusion Reaction*