The Carpentier-Edwards pericardial aortic valve. Ten-year results

J Thorac Cardiovasc Surg. 1995 Sep;110(3):651-62. doi: 10.1016/S0022-5223(95)70096-X.

Abstract

To evaluate the function of the Carpentier-Edwards pericardial valve in the aortic position, we analyzed the results of 310 aortic valve replacements performed between 1982 and 1985. Mean age was 64.2 +/- 10.8 years (range 22 to 95 years); 190 patients (61.3%) were male patients. There were 18 hospital deaths (5.8%), and none were valve related. Follow-up of the 292 survivors was 100% complete at a mean of 7.8 +/- 2.9 years; 2290 patient-years of follow-up were available for analysis. There were 133 late deaths (45.5%). Actuarial survivals at 5 and 10 years were 82.5% and 45.9%, respectively. The 10-year actuarial freedom from events was 88.7% +/- 2.1% for thromboembolism, 90.9% +/- 1.8% for hemorrhage, 94.3% +/- 1.6% for endocarditis, and 91.2% +/- 2.6% for structural deterioration. The 153 hospital survivors 65 years of age or older had an extremely low incidence of structural valve deterioration, with only four explants and 95.5% actuarial freedom from explantation at 10 years, and a linearized rate of 0.3 +/- 0.2 per patient-year compared with 88.6% and 0.7 +/- 0.2 for patients younger than 65 years of age. Twelve valves were explanted for structural deterioration. Of these, 11 (93%) had leaflet calcification causing stenosis and one had a wear-related leaflet tear. The Carpentier-Edwards pericardial valve has a low incidence of valve-related complications. The freedom from structural valve deterioration is low at 10 years, particularly in patients 65 years of age and older.

MeSH terms

  • Actuarial Analysis
  • Adult
  • Aged
  • Aged, 80 and over
  • Aortic Valve / surgery
  • Bioprosthesis*
  • Disease-Free Survival
  • Endocarditis / etiology
  • Female
  • Follow-Up Studies
  • Heart Valve Prosthesis* / adverse effects
  • Heart Valve Prosthesis* / mortality
  • Hemorrhage / etiology
  • Humans
  • Logistic Models
  • Male
  • Middle Aged
  • Multivariate Analysis
  • Pericardium
  • Proportional Hazards Models
  • Prosthesis Failure
  • Survival Rate
  • Thromboembolism / etiology