Impact of soft tissue on in vivo accuracy of bone mineral measurements in the spine, hip, and forearm: a human cadaver study

J Bone Miner Res. 1995 Jun;10(6):868-73. doi: 10.1002/jbmr.5650100607.

Abstract

The impact of soft tissue in vivo on accuracy of bone mineral density (BMD) measurements of the spine and hip by dual energy X-ray absorptiometry and of the forearm by single photon absorptiometry was assessed by use of 14 human cadavers. The in vivo accuracy errors (SEE%) were: forearm 3-5%, anteroposterior spine 5.3%, lateral spine 10-12%, and femoral greater trochanter, neck, total, intertrochanteric, and Ward's triangle 3%, 6.5%, 6.7%, 8%, and 11-13%, respectively. Except from the lateral spine and the greater trochanter, the slopes of the linear regressions of in vivo BMD against in vitro BMD were not significantly different from 1 (p > 0.05). The calculated random accuracy error of BMD measurements due to fat inhomogeneity was estimated to 3-4% for the anteroposterior spine and 9-14% for the lateral spine (from abdominal computed tomography in 26 healthy women). In conclusion, acceptable accuracy errors below 6-7% (of soft tissue in vivo) of BMD measurements were obtained in the anteroposterior spine, the forearm, and the neck, greater trochanter, and total proximal femur.

Publication types

  • Comparative Study

MeSH terms

  • Absorptiometry, Photon
  • Aged
  • Aged, 80 and over
  • Bone Density / physiology*
  • Cadaver
  • Connective Tissue / metabolism*
  • Connective Tissue / physiology
  • Female
  • Forearm / physiology
  • Hip / physiology
  • Humans
  • In Vitro Techniques
  • Linear Models
  • Male
  • Middle Aged
  • Reproducibility of Results
  • Spine / physiology
  • Tomography, X-Ray Computed