Abstract

A newly devised, very accurate ("definitive") method for the determination of trace amounts of cobalt in biological materials was validated by the analysis of several certified reference materials. The method is based on a combination of neutron activation and selective and quantitative postirradiation isolation of radiocobalt from practically all other radionuclides by ion-exchange and extraction chromatography followed by gamma-ray spectrometric measurement. The significance of criteria that should be fulfilled in order to accept a given result as obtained by the "definitive method" is emphasized. In view of the demonstrated very good accuracy of the method, it is suggested that our values for cobalt content in those reference materials in which it was originally not certified (SRM 1570 spinach, SRM 1571 orchard leaves, SRM 1577 bovine liver, and Czechoslovak bovine liver 12-02-01) might be used as provisional certified values.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Animals
  • Cattle
  • Chromatography, Ion Exchange
  • Cobalt / analysis*
  • Liver / chemistry
  • Plant Leaves / chemistry
  • Quality Control
  • Reference Standards
  • Spectrometry, Gamma
  • Spinacia oleracea / chemistry

Substances

  • Cobalt