We describe the use of electrical impedance tomography (EIT) for non-invasive thermal imaging in conjunction with a clinical treatment of a superficial scalp lesion utilizing a spiral microstrip antenna. This is our first reported use of EIT with a clinical hyperthermia treatment and perhaps the first world-wide. The thermal measurements recorded during treatment compare favourably with the images reconstructed from impedance data gathered during heating. A linear relation, measured in phantom material, between the change in temperature with the change in reconstructed impedance was assumed. The average discrepancy between the measured temperature changes with the temperatures reconstructed from the impedance changes was 1.4 degrees C, with the maximum being 8.9 degrees C. These preliminary data suggest that impedance changes can be measured during hyperthermia delivery and temperature estimates based on these observed changes are possible in the clinical setting. These findings also point to the complex, yet critical nature of the impedance versus temperature relationship for tissue in vivo. The reconstructed thermal images may provide complementary information about the overall thermal damage imposed during heating. Based on this initial clinical experience we feel that EIT has great potential as a viable clinical aid in imaging the temperature changes imposed during hyperthermia.