Tagged MR imaging in a deforming phantom: photographic validation

Radiology. 1994 Mar;190(3):765-9. doi: 10.1148/radiology.190.3.8115625.

Abstract

Purpose: To validate cine magnetic resonance (MR) image tagging measurements of a deforming object by means of a precise photographic method.

Materials and methods: A piece of silicone rubber that acted as a phantom was stretched in a cyclical fashion inside a plastic clamp driven by a respirator pump. Deformation as a function of time was measured with a rapid gradient-echo cine tagging sequence and with sequential stroboscopic photographs. Deformations from 1.0 to 1.2 (0% to 20% stretch) in the readout direction were measured over a 7-cm region of the phantom, which had a maximum standard error of +/- 0.001 with photography and a maximum standard error of +/- 0.003 with MR imaging.

Results: The deformation versus time values measured with MR imaging had a standard error of 0.002 about a straight line fit to the photographic deformation versus time data. These results demonstrate that the MR imaging deformation estimates were accurate and precise.

Conclusion: The validated tagging method can now be used to evaluate MR imaging motion estimation techniques.

Publication types

  • Research Support, Non-U.S. Gov't
  • Research Support, U.S. Gov't, P.H.S.

MeSH terms

  • Artifacts*
  • Humans
  • Image Processing, Computer-Assisted
  • Magnetic Resonance Imaging / methods*
  • Magnetic Resonance Imaging / standards
  • Models, Structural
  • Motion
  • Photography
  • Silicone Elastomers