The method outlined in this paper utilizes internal standardization and is simple, reliable, and sensitive for the determination of fluvastatin in plasma. Fluvastatin sodium and the internal standard are extracted from buffered plasma into methyl tert.-butyl ether, followed by evaporation of an aliquot of the organic phase. After reconstitution of the dried sample into a small volume of mobile phase (methanol-13 mM tetrabutylammonium fluoride, 3:2, v/v), the sample is chromatographed on an LC-18 column thermostated at 50 degrees C. Fluorescence detection (excitation at 305 nm and emission at 380 nm) is used to monitor both fluvastatin (free acid) and the internal standard. The method can accurately detect 1 ng/ml fluvastatin using a 1.0-ml plasma sample. The precision and reproducibility over the linear range of the method are 5.57 and 7.32%, respectively. This method has been used to measure fluvastatin plasma concentrations in support of bioavailability/pharmacokinetic studies with no indication of interference.