Adaptive finite element modeling of long-term polyethylene wear in total hip arthroplasty

J Orthop Res. 1996 Jul;14(4):668-75. doi: 10.1002/jor.1100140424.

Abstract

Adaptive remeshing capability was added to an existing sliding-distance-coupled finite element model of polyethylene wear in total hip arthroplasty. This augmentation allowed earlier postoperative wear simulation to be extended to the clinically more significant long-term regimen (as long as 20 years). Loads and femoral head excursions were taken from a physically validated gait analysis model of a patient with an instrumented total hip replacement. For otherwise identical 22, 28, and 32 mm components, the least volumetric wear but the most linear wear occurred for the 22 mm head. When the polyethylene thickness in a 22 mm component was reduced to the same as that in a 32 mm component, the volumetric wear rate for the 22 mm component was still much less than that for the larger component, indicating that sliding distance (head size), rather than polyethylene liner thickness, was primarily responsible for the difference in rates. A "28 mm" series, for which head sizes were varied across the range of currently accepted industrial tolerances, showed that although initial wear rates were greatest for the least congruent articulations, the long-term volumetric wear was nearly the same, regardless of initial clearance.

Publication types

  • Research Support, U.S. Gov't, P.H.S.

MeSH terms

  • Arthroplasty*
  • Femur Head
  • Hip Prosthesis / instrumentation*
  • Humans
  • Models, Biological*
  • Polyethylenes
  • Time Factors

Substances

  • Polyethylenes