For defibrillation between right ventricular and retropectoral patch electrodes using truncated exponential pulses, the stored energy defibrillation threshold (DFT) is lower for short pulses from small 60-microF capacitors than for conventional pulses from 120-microF capacitors, but 60-microF pulses frequently require higher voltages than are currently used. The goal of this study was to determine if DFT could be reduced by intermediate size 90-microF capacitors. This study compared biphasic waveform DFTs for 120 microF-65% tilt pulses, 90 microF-65% tilt pulses, and 90 microF-50% tilt pulses in 20 patients at defibrillator implantation. The 90 microF-50% tilt pulses were selected because their duration is half that of 120 microF-65% tilt pulses. The stored energy DFT for 90 microF-50% tilt pulses (9.1 +/- 4.3 J) was less than both the DFT for 120 microF-65% tilt pulses (12.0 +/- 5.5 J, P < 0.005) and the DFT for 90 microF-65% tilt pulses (11.6 +/- 5.8 J, P < 0.005). There was no significant difference between the latter two values. The voltage DFTs for 90 microF-50% pulses (436 +/- 113 V) and 120 microF-65% tilt pulses (436 +/- 104 V) were not statistically different; the voltage DFT for 90 microF-65% tilt pulses was higher than for either of the other two pulses (490 +/- 131, P < 0.005). The DFT was 20 J or greater in three patients for both 120 microF-65% tilt pulses and 90 microF-65% tilt pulses, but it was 16 J or less in all patients for 90 microF-50% tilt pulses. When pathways were dichotomized by the median resistance of 71 omega, 90 microF-50% tilt pulses significantly reduced DFTs compared to 120 microF-65% tilt pulses for higher resistance pathways (9.2 +/- 4.0 J vs 13.0 +/- 6.2 J, P = 0.002), but not lower resistance pathways (9.0 +/- 4.8 J vs 10.9 +/- 4.6 J, P = NS). For the electrode configuration tested, biphasic 90 microF-50% tilt pulses reduce stored energy DFT in comparison with 120 microF-65% tilt pulses without increasing voltage DFT. However, 90 microF-65% tilt pulses provide no benefit.