Diagnosis of premature rupture of the membranes in preterm patients, using an improved AFP kit: comparison with ROM-check and/or nitrazine test

Eur J Obstet Gynecol Reprod Biol. 1996 Nov;69(2):77-82. doi: 10.1016/0301-2115(95)02519-7.

Abstract

Objective: We developed a new kit for detecting AFP in leaked amniotic fluid. Later, we developed an improved AFP kit utilizing the same anti alpha-fetoprotein (AFP) monoclonal antibody. In this study, we evaluate the clinical usefulness of this improved kit in the diagnosis of preterm premature rupture of membranes (PROM).

Methods: We compared this improved AFP test with the ROM-check and/or the nitrazine tests in 46 preterm patients.

Results: The ROM-check and nitrazine tests showed a diagnostic accuracy of 89.1 and 87.0%, respectively, compared with 95.7% with the improved AFP test. The sensitivity of the improved AFP test on cervical samples was significantly higher than that of the nitrazine test on vaginal samples (P < 0.05). The reaction time with the improved AFP kit test is 90 s.

Conclusion: This study has confirmed a great clinical utility of the improved AFP test kit as a method of PROM diagnosis.

Publication types

  • Comparative Study

MeSH terms

  • Amniotic Fluid / chemistry*
  • Amniotic Fluid / immunology
  • Antibodies, Monoclonal
  • False Positive Reactions
  • Female
  • Fetal Membranes, Premature Rupture / diagnosis*
  • Humans
  • Immunosorbent Techniques
  • Pregnancy
  • Prenatal Diagnosis / methods*
  • Prospective Studies
  • Reagent Kits, Diagnostic
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Time Factors
  • alpha-Fetoproteins / analysis*
  • alpha-Fetoproteins / immunology

Substances

  • Antibodies, Monoclonal
  • Reagent Kits, Diagnostic
  • alpha-Fetoproteins