European multicentre evaluation of a commercial system for identification of methicillin-resistant Staphylococcus aureus

Eur J Clin Microbiol Infect Dis. 1996 Sep;15(9):747-9. doi: 10.1007/BF01691964.

Abstract

A commercial system for the rapid detection of methicillin-resistant Staphylococcus aureus, the BBL Crystal MRSA test (C-MRSA ID; Becton Dickinson, USA), was evaluated prospectively and compared with a polymerase chain reaction test for the presence of the mecA gene. Ten European centres tested a total of 676 isolates of Staphylococcus aureus from blood cultures. The system correctly identified 661 (97.8%) isolates within 4 h. All but three mecA gene-negative isolates (99.4% specificity) yielded a negative C-MRSA ID reaction, and 158 of 170 mecA gene-positive isolates were accurately detected (92.9% sensitivity). After repeated testing of discrepant results, sensitivity and specificity increased to 99% and 100%, respectively.

MeSH terms

  • Bacteriological Techniques*
  • Evaluation Studies as Topic
  • Humans
  • Methicillin Resistance*
  • Microbial Sensitivity Tests
  • Multicenter Studies as Topic
  • Polymerase Chain Reaction*
  • Predictive Value of Tests
  • Prospective Studies
  • Sensitivity and Specificity
  • Staphylococcal Infections / diagnosis
  • Staphylococcal Infections / drug therapy*
  • Staphylococcus aureus / drug effects
  • Staphylococcus aureus / isolation & purification*