Quality control phantom for digital chest radiography

Radiology. 1997 Jan;202(1):111-6. doi: 10.1148/radiology.202.1.8988199.

Abstract

Purpose: To develop and test a chest phantom for routine quality control testing of digital radiography systems.

Materials and methods: The phantom was constructed from sheets of copper, aluminum, and acrylic, which were cut and arranged to yield a radiographic projection resembling that of a human thorax. Regional test objects allowed quantitative assessment of optical density, contrast detail, and spatial resolution. Validation tests were performed to assess image stability in a stable imaging environment and sensitivity to changes in image quality when they occur.

Results: The phantom yielded consistent pseudoclinical images when used in a routine quality control program and facilitated detection of simulated problems that were induced in imaging system performance.

Conclusion: The chest phantom enables quantitative, full-system testing of digital radiography system as they are used clinically for chest radiography.

MeSH terms

  • Humans
  • Phantoms, Imaging*
  • Quality Control
  • Radiographic Image Enhancement / standards*
  • Radiography, Thoracic / standards*