Low level determination of a novel 4-azasteroid and its carboxylic acid metabolite in human plasma and semen using high-performance liquid chromatography with atmospheric pressure chemical ionization tandem mass spectrometry

J Chromatogr B Biomed Sci Appl. 1997 May 23;693(1):117-29. doi: 10.1016/s0378-4347(97)00047-9.

Abstract

Compound I (4.7beta-dimethyl-4-azacholestan-3-one, MK-0386) is a potent 5alpha-reductase type 1 (5alphaR1) inhibitor. Sensitive (0.2 ng/ml), specific and separate assays have been developed and validated for the analysis of I and its carboxylic acid metabolite (II) in human semen and plasma based on high-performance liquid chromatography (HPLC) with tandem mass spectrometric (MS-MS) detection. After liquid-liquid extraction of the analytes from biological matrix, the extracts were chromatographed on a short (50 mm) analytical column during analysis of I, and on a longer (150 mm) column with a weaker mobile phase during the analysis of II. This additional chromatographic separation was required to separate II from a secondary metabolite present in post-dose plasma samples interfering with the quantification of II. The MS-MS detection was performed on a Sciex API III Plus tandem mass spectrometer using the heated nebulizer probe. Monitoring the parent-->product ion combinations of m/z 416-->114 and 404-->114, in the multiple reaction monitoring (MRM) mode, after chromatographic separation, allowed quantification of both analytes. The standard curve in plasma was linear in the concentration range of 0.2 to 200 ng/ml for both I and II with correlation coefficients greater than 0.99 and coefficients of variation of less than 15% for replicate (n=5) analysis at all concentrations within the standard curve range. For the semen assay the linear range for determination of I was from 0.2 to 50 ng/ml. These assays were applied to support a number of clinical studies with I and their validity and long-term performance was confirmed during analyses of clinical samples from these studies. The need for careful assessment of the specificity of MS-MS assays in post-dose biological fluid samples in the presence of metabolites was emphasized.

MeSH terms

  • 5-alpha Reductase Inhibitors*
  • Azasteroids / analysis
  • Azasteroids / blood*
  • Azasteroids / metabolism
  • Chromatography, High Pressure Liquid
  • Enzyme Inhibitors / analysis
  • Enzyme Inhibitors / blood*
  • Enzyme Inhibitors / metabolism
  • Humans
  • Male
  • Mass Spectrometry
  • Semen / chemistry*
  • Sensitivity and Specificity