In vivo morphology of woven, collagen-sealed Dacron prostheses in the thoracic aorta

Ann Thorac Surg. 1997 Oct;64(4):1096-8. doi: 10.1016/s0003-4975(97)00802-3.

Abstract

Background: Long-term changes in knitted Dacron grafts inserted into the infrarenal aorta have been addressed by a number of studies indicating their potential for postoperative dilatation. In contrast, the behavior of woven, collagen-presealed, double-velour Dacron grafts used to replace the thoracic aorta is not known.

Methods: Forty-five patients were examined at a mean of 32.4 +/- 14.8 months after insertion of woven, collagen-coated, Dacron double-velour prostheses (Meadox woven with Hemashield, Meadox, Oakland, NJ) in the thoracic position under highly standardized conditions using spiral computed tomography.

Results: Compared with a manufactured diameter of 26 mm, all grafts showed an increase of 1 to 5 mm (mean, 3.0 +/- 1.2 mm [11.6% +/- 4.4%]; p < 0.0001) with greater enlargement of the ascending than of the descending aortic portions (p = not significant). A further statistically significant progressive dilatation failed to occur. Degenerative changes, including false aneurysm formation, could be excluded.

Conclusions: Woven, collagen-coated Dacron prostheses are considered a safe replacement material for the thoracic aorta.

MeSH terms

  • Aorta, Thoracic / diagnostic imaging
  • Aorta, Thoracic / surgery*
  • Aortic Aneurysm, Thoracic
  • Aortic Dissection
  • Blood Vessel Prosthesis*
  • Collagen*
  • Disease Progression
  • Female
  • Follow-Up Studies
  • Humans
  • Image Processing, Computer-Assisted
  • Male
  • Middle Aged
  • Polyesters
  • Polyethylene Terephthalates*
  • Prosthesis Design
  • Radiography
  • Textiles

Substances

  • Polyesters
  • Polyethylene Terephthalates
  • velour
  • Collagen