[Dosimetry in conventional and spiral CT in 8 spiral CT scanners of the latest generation]

Rofo. 1998 Jun;168(6):562-6. doi: 10.1055/s-2007-1015281.
[Article in German]

Abstract

Purpose: Measurement of scanner-dependent differences in PCTDI for conventional and spiral CT and evaluation of PCTDI/100 mAs in eight spiral CT scanners of the latest generation.

Methods: Central PCTDI was evaluated in a PMMA phantom for different scan lengths of conventional and spiral CT in four scanners. Central and peripheral PCTDI/100 mAs, average dose and the quotient between peripheral and central PCTDI/100 mAs were measured with a dedicated 10 cm ionisation chamber positioned within a PMMA phantom using the parameters of a standardised liver protocol and a high-resolution inner ear protocol for eight spiral CT scanners.

Results: Depending on scanner type and scan length PCTDI for the spiral mode is in a range between 11.3% higher and 4.3% lower than that for the conventional mode. PCTDI/100 mAs varies considerably between different CT scanners. For the standardised liver protocol PCTDI/100 mAs varies up to a factor of 2, for the inner ear protocol up to a factor of 4.

Conclusion: Even in comparable modern CT scanners there are considerable protocol and scanner-dependent differences in PCTDI/100 mAs. The values for his own scanner should be known to each CT user so that he can develop a dose-optimised, patient and problem-orientated examination strategy.

Publication types

  • Comparative Study
  • Multicenter Study

MeSH terms

  • Ear, Inner / diagnostic imaging
  • Humans
  • Liver / diagnostic imaging
  • Models, Theoretical
  • Phantoms, Imaging
  • Radiation Dosage*
  • Tomography, X-Ray Computed / instrumentation
  • Tomography, X-Ray Computed / methods*