Effects of dose reduction on the detectability of standardized radiolucent lesions in digital panoramic radiography

Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 1998 Aug;86(2):227-33. doi: 10.1016/s1079-2104(98)90130-5.

Abstract

Dose reduction in digital panoramic radiography was studied. Intentional underexposure was performed with the Orthophos DS while six different human mandibles were radiographed. Exposure settings were 69 kV/15 mA (standard), 64 kV/16 mA, and 60 kV/16 mA. Standardized spherical defects, each either 1 or 1.25 mm in diameter, were simulated in 288 of 432 images, and seven observers decided whether defects were present or not. Areas under the receiver operating characteristics curves were calculated. They showed no significant differences in the detectability of the 1-mm defect at 69, 64, or 60 kV. For the 1.25-mm defect, no difference was found between the 69 and 60 kV images, but a statistically significant different detectability was found for 64 kV images in comparison with both 69 and 60 kV images. A dose reduction of up to 43% was ascertained with a Pedo-RT-Humanoid phantom when panoramic radiography was performed at 60 kV/16 mA. The conclusion is that with the Orthophos DS, it seems possible to reduce the dose rate of x-rays without loss of diagnostic quality in the case of radiolucent changes.

Publication types

  • Comparative Study

MeSH terms

  • Fluorides
  • Humans
  • Image Processing, Computer-Assisted
  • Lithium Compounds
  • Mandibular Diseases / diagnostic imaging
  • Observer Variation
  • Phantoms, Imaging
  • ROC Curve
  • Radiation Dosage*
  • Radiographic Image Enhancement / instrumentation
  • Radiographic Image Enhancement / methods
  • Radiography, Dental, Digital / instrumentation
  • Radiography, Dental, Digital / methods*
  • Radiography, Panoramic / instrumentation
  • Radiography, Panoramic / methods*
  • Thermoluminescent Dosimetry / instrumentation

Substances

  • Lithium Compounds
  • lithium fluoride
  • Fluorides