In an experimental study (in vitro and in vivo) we evaluated the efficacy of various biopsy needles/devices for breast biopsy. In vitro, biopsies of five human cadaveric breast specimens were performed using 33 different needles/devices ranging from 14 to 20-gauge. Of these 33 needles/devices, 22 optimally performing needles were selected for the in vivo study. In the clinical part of the study, 44 breast lesions were randomly biopsied with each of the 22 needles/devices under stereotactic guidance. Tissue specimens were analysed quantitatively and qualitatively. Several automatic long-throw guns (Acecut, Asap, Biopty, Magnum) obtained greater tissue areas and had a better histopathologic score than the conventional type of a side-notch needle like Trucut, an aspiration needle like Surecut, or an end-cut needle like Autovac. The automatic long-throw guns performed better than the short-throw Monopty gun. Regardless of needle size (14-20-gauge), breast biopsies should be routinely performed with automated long-throw side-notch guns (Acecut, Asap, Biopty, Magnum).