Programming optimal atrioventricular delay in dual chamber pacing using peak endocardial acceleration: comparison with a standard echocardiographic procedure.
Dupuis JM, Kobeissi A, Vitali L, Gaggini G, Merheb M, Rouleau F, Leftheriotis G, Ritter P, Victor J.
Dupuis JM, et al. Among authors: ritter p.
Pacing Clin Electrophysiol. 2003 Jan;26(1P2):210-3. doi: 10.1046/j.1460-9592.2003.00018.x.
Pacing Clin Electrophysiol. 2003.
PMID: 12687814
The OAVD was also evaluated for each patient using the Ritter echocardiographic method. Good sigmoid fit was obtained in 13 of 15 patients. The mean OAVD obtained by the PEA sigmoid algorithm was 146.9 +/- 32.1 ms, and the corresponding result obtained by echocardiography …