Introduction: Linked color imaging (LCI) and blue laser imaging-bright (BLI-b) improve the visibility of gastrointestinal lesions. In this multicenter study, we compared the effects of LCI and BLI-b on the visibility of flat polyps with visibility scores and color difference (CD) values, including fast-withdrawal and large-monitor observation.
Methods: We recorded 120 videos of 40 consecutive flat polyps (2-20 mm), adenoma, and sessile serrated adenoma and polyp (SSA/P), using white light imaging (WLI), BLI-b, and LCI from July 2017 to December 2017. All videos were evaluated by eight endoscopists according to a published polyp visibility score of 4 (excellent) to 1 (poor). Additionally, 1.5 ×faster and 1.7 ×sized videos were evaluated. Moreover, we calculated the CD values for each polyp in three modes.
Results: The mean LCI scores (3.1 ± 0.9) were significantly higher than the WLI scores (2.5 ± 1.0, p < 0.001) but not significantly higher than the BLI-b scores (3.0 ± 1.0). The scores of faster videos on LCI (3.0 ± 1.1) were significantly higher than WLI (2.0 ± 1.0, p < 0.001) and BLI-b (2.8 ± 1.1, p = 0.03). The scores of larger-sized videos on LCI were not significantly higher than those of WLI or BLI-b. The CD value of LCI (18.0 ± 7.7) was higher than that of WLI (11.7 ± 7.0, p < 0.001), but was not significantly higher than that of BLI-b (16.6 ± 9.6). The CD value of LCI was significantly higher than that of BLI-b for adenoma, but the CD value of BLI-b was significantly higher than that of LCI for SSA/P.
Conclusions: The superiority of LCI to BLI-b was proven for the visibility of adenoma and fast observation.
Keywords: Blue laser imaging; Colonic polyps; Colonoscopy; LASER; Linked color imaging.