A new irradiation system for lung cancer combining linear accelerator, computed tomography, patient self-breath-holding, and patient-directed beam-control without respiratory monitoring devices

Int J Radiat Oncol Biol Phys. 2003 May 1;56(1):14-20. doi: 10.1016/s0360-3016(02)04414-0.

Abstract

Purpose: To introduce and assess a new irradiation technique for lung cancer that utilizes a linear accelerator and computed tomography (CT) scanner combination, along with a novel switching mechanism, which enables patients to synchronize the duration of irradiation with self-breath-holding without respiratory monitoring devices.

Materials and methods: A newly developed treatment unit, a linear accelerator combined with a CT scanner (CT-linac), was used for irradiation. A novel switching mechanism, connected directly to the console of the linear accelerator, enabled the patient to control the radiation beam to correspond with the duration of self-breath-holding during a session determined by a radiation technologist. Twenty patients with lung cancer were enrolled in this study. All patients were instructed in the technique of breath-holding during the inspiration phase using visualization of respiratory motion through fluoroscopy as a teaching aid. CT scans under patients' self-breath-holding were repeated three times, and differences in tumor position on CT images were measured. The reproducibility of tumor position was visually evaluated on electronic portal images (EPI).

Results: Mean maximum differences in tumor position under patients' self-breath-holding were 2.2 mm in the cranial-caudal direction, 1.4 mm in the anterior-posterior direction, and 1.3 mm in the right-left direction. Switching of the radiation beam was delayed less than 0.1 s behind patient switching. EPIs were used to determine that reproducibility of tumor position was satisfactorily accurate.

Conclusions: The reproducibility of tumor position, during patient self-breath-holding synchronized with patient-initiated radiation and without a respiratory monitoring device, was sufficiently accurate. This novel irradiation technique for lung tumors using a combination CT-linac offers reduced PTV, sufficient reproducibility, and decreased duration of treatment.

Publication types

  • Evaluation Study

MeSH terms

  • Aged
  • Aged, 80 and over
  • Artifacts
  • Computer Systems
  • Equipment Design
  • Female
  • Humans
  • Inhalation
  • Lung Neoplasms / diagnostic imaging
  • Lung Neoplasms / radiotherapy*
  • Male
  • Man-Machine Systems*
  • Middle Aged
  • Monitoring, Physiologic
  • Particle Accelerators*
  • Radiotherapy Planning, Computer-Assisted
  • Radiotherapy, High-Energy / instrumentation*
  • Radiotherapy, High-Energy / methods
  • Reproducibility of Results
  • Respiration*
  • Tomography, X-Ray Computed / instrumentation*
  • Volition