Characteristics of EPI-hNE4 aerosol: a new elastase inhibitor for treatment of cystic fibrosis

J Aerosol Med. 2003 Summer;16(2):121-9. doi: 10.1089/089426803321919889.

Abstract

The purpose of this study was to define nebulization conditions providing delivery of aerosols of EPI-hNE4, an inhibitor of human neutrophil elastase (HNE). EPI-hNE4 was nebulized with Pari LC Star and tested at three concentrations (2.5, 5, and 10 mg/mL). The inhaled mass was measured over 15 min. Particle size distribution was measured by cascade impaction. The effect was also tested of mixing EPI-hNE4 with a (99m)Tc human serum albumin (HSA) tracer on the aerodynamic properties of the aerosol. The inhibitory activity of EPI-hNE4 after nebulization was assessed on purified HNE. The inhaled mass was 32.3 +/- 3.5% (mean +/- SD) after 10 min and 44.2 +/- 3.8% (mean +/- SD) after 15 min. Mass median aerodynamic diameter ranged between 1.2 and 1.8 microm. The (99m)Tc HSA EPI-hNE4 aerosol was similar in terms of particle size distribution (y = 1.0338x - 0.003, r = 0.83). (99m)Tc activity was predictive of EPI-hNE4 mass distribution (y = 1.0278x - 1.6991, r = 0.89). The inhibitory capacity of aerosolized samples remained unchanged after up to 10 min of nebulization. EPI-hNE4 can be nebulized efficiently without decrease in its activity. Mixing this inhibitor with (99m)Tc HSA should allow quantification of its deposition in CF patients.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Administration, Inhalation
  • Aerosols
  • Cystic Fibrosis / drug therapy*
  • Equipment Design
  • Humans
  • Linear Models
  • Nebulizers and Vaporizers*
  • Particle Size
  • Proteins / administration & dosage*
  • Serpins

Substances

  • Aerosols
  • Proteins
  • Serpins
  • SERPINB1 protein, human