Dual-chamber versus single-chamber detection enhancements for implantable defibrillator rhythm diagnosis: the detect supraventricular tachycardia study

Circulation. 2006 Jun 27;113(25):2871-9. doi: 10.1161/CIRCULATIONAHA.105.594531. Epub 2006 Jun 12.

Abstract

Background: Delivery of inappropriate shocks caused by misdetection of supraventricular tachycardia (SVT) remains a substantial complication of implanted cardioverter/defibrillator (ICD) therapy. Whether use of optimally programmed dual-chamber ICDs lowers this risk compared with that in single-chamber ICDs is not clear.

Methods and results: Subjects with a clinical indication for ICD (n=400) at 27 participating centers received dual-chamber ICDs and were randomly assigned to strictly defined optimal single- or dual-chamber detection in a single-blind manner. Programming minimized ventricular pacing. The primary end point was the proportion of SVT episodes inappropriately detected from the time of programming until crossover or end of study. On a per-episode basis, 42% of the episodes in the single-chamber arm and 69% of the episodes in the dual-chamber arm were due to SVT. Mortality (3.5% in both groups) and early study withdrawal (14% single-chamber, 11% dual-chamber) were similar in both groups. The rate of inappropriate detection of SVT was 39.5% in the single-chamber detection arm compared with 30.9% in the dual-chamber arm. The odds of inappropriate detection were decreased by almost half with the use of the dual-chamber detection enhancements (odds ratio, 0.53; 95% confidence interval, 0.30 to 0.94; P=0.03).

Conclusions: Dual-chamber ICDs, programmed to optimize detection enhancements and to minimize ventricular pacing, significantly decrease inappropriate detection.

Publication types

  • Comparative Study
  • Multicenter Study
  • Randomized Controlled Trial
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Aged
  • Arrhythmias, Cardiac / diagnosis*
  • Arrhythmias, Cardiac / physiopathology
  • Arrhythmias, Cardiac / therapy
  • Atrioventricular Node / physiology
  • Cardiac Pacing, Artificial / methods
  • Confidence Intervals
  • Cross-Over Studies
  • Defibrillators, Implantable / adverse effects
  • Defibrillators, Implantable / standards*
  • Diagnosis, Differential
  • Electric Countershock / instrumentation*
  • Electric Countershock / methods
  • Electrocardiography
  • Electrophysiologic Techniques, Cardiac / methods*
  • Equipment Design
  • Equipment Failure
  • Female
  • Heart Rate / physiology*
  • Humans
  • Male
  • Middle Aged
  • Sensitivity and Specificity
  • Single-Blind Method
  • Tachycardia, Supraventricular / diagnosis*
  • Tachycardia, Supraventricular / physiopathology
  • Tachycardia, Supraventricular / therapy
  • Tachycardia, Ventricular / diagnosis
  • Tachycardia, Ventricular / physiopathology
  • Tachycardia, Ventricular / therapy
  • Time Factors
  • Ventricular Fibrillation / diagnosis
  • Ventricular Fibrillation / physiopathology
  • Ventricular Fibrillation / therapy