Reliability of the IOLMaster in axial length evaluation in silicone oil-filled eyes

Eye (Lond). 2007 Jul;21(7):909-11. doi: 10.1038/sj.eye.6702452. Epub 2006 Jul 28.

Abstract

Purpose: To assess the reliability of IOLMaster in axial length (AL) measurement in phakic silicone oil-filled vs pseudophakic saline-filled eyes.

Methods: Ten eyes of 10 patients, vitrectomized with silicone oil tamponade and scleral buckled with significant lens opacity were enrolled. Optical biometry with IOLMaster (Carl Zeiss Meditec AG, Germany) was performed 1 day before and 1 week after silicone oil removal and phacoemulsification with artificial intraocular lens (IOL) implantation in order to assess changes in AL measurements.

Results: Mean AL was 26.16+/-1.23 mm (range 24.64-28.8 mm) and 26.27+/-1.46 mm (range 25.26-29.6 mm), respectively, the day before and 1 week after silicone oil removal and cataract surgery, and the difference was not statistically significant (P=0.2).

Conclusions: Presence vs absence of silicone oil tamponade as well as phakic vs pseudophakic status in buckled and vitrectomized eyes did not influence the AL measurement by means of no-contact optical biometry, suggesting that such eyes might be candidate for silicone oil removal and cataract surgery at one time.

Publication types

  • Evaluation Study

MeSH terms

  • Aged
  • Aphakia, Postcataract / physiopathology
  • Biometry / instrumentation*
  • Biometry / methods
  • Drainage
  • Female
  • Humans
  • Lens Implantation, Intraocular
  • Male
  • Middle Aged
  • Phacoemulsification*
  • Prospective Studies
  • Pseudophakia / physiopathology
  • Refraction, Ocular
  • Retinal Detachment / surgery
  • Scleral Buckling
  • Silicone Oils*
  • Vitrectomy
  • Vitreous Body

Substances

  • Silicone Oils