Strategy for safe performance of extrathoracic magnetic resonance imaging at 1.5 tesla in the presence of cardiac pacemakers in non-pacemaker-dependent patients: a prospective study with 115 examinations

Circulation. 2006 Sep 19;114(12):1285-92. doi: 10.1161/CIRCULATIONAHA.105.597013. Epub 2006 Sep 11.

Abstract

Background: The purpose of the present study was to evaluate a strategy for safe performance of extrathoracic magnetic resonance imaging (MRI) in non-pacemaker-dependent patients with cardiac pacemakers.

Methods and results: Inclusion criteria were presence of a cardiac pacemaker and urgent clinical need for an MRI examination. Pacemaker-dependent patients and those requiring examinations of the thoracic region were excluded. The study group consisted of 82 pacemaker patients who underwent a total of 115 MRI examinations at 1.5T. To minimize radiofrequency-related lead heating, the specific absorption rate was limited to 1.5 W/kg. All pacemakers were reprogrammed before MRI: If heart rate was <60 bpm, the asynchronous mode was programmed to avoid magnetic resonance (MR)-induced inhibition; if heart rate was >60 bpm, sense-only mode was used to avoid MR-induced competitive pacing and potential proarrhythmia. Patients were monitored with ECG and pulse oximetry. All pacemakers were interrogated immediately before and after the MRI examination and after 3 months, including measurement of pacing capture threshold (PCT) and serum troponin I levels. All MR examinations were completed safely. Inhibition of pacemaker output or induction of arrhythmias was not observed. PCT increased significantly from pre- to post-MRI (P=0.017). In 2 of 195 leads, an increase in PCT was only detected at follow-up. In 4 of 114 examinations, troponin increased from a normal baseline value to above normal after MRI, and in 1 case (troponin pre-MRI 0.02 ng/mL, post-MRI 0.16 ng/mL), this increase was associated with a significant increase in PCT.

Conclusions: Extrathoracic MRI of non-pacemaker-dependent patients can be performed with an acceptable risk-benefit ratio under controlled conditions and by taking both MR- and pacemaker-related precautions.

Publication types

  • Clinical Trial
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Abdomen / pathology
  • Adolescent
  • Adult
  • Aged
  • Aged, 80 and over
  • Arrhythmias, Cardiac / etiology
  • Arrhythmias, Cardiac / physiopathology
  • Arrhythmias, Cardiac / prevention & control*
  • Brain / pathology
  • Child
  • Child, Preschool
  • Electrocardiography
  • Equipment Safety / instrumentation*
  • Equipment Safety / methods
  • Female
  • Heart Rate / physiology
  • Heating
  • Humans
  • Magnetic Resonance Imaging / adverse effects*
  • Magnetic Resonance Imaging / methods*
  • Male
  • Middle Aged
  • Pacemaker, Artificial*
  • Pelvis / pathology
  • Prospective Studies
  • Radio Waves / adverse effects
  • Risk Assessment
  • Troponin I / blood

Substances

  • Troponin I