Two-channel gastric pacing with a novel implantable gastric pacemaker accelerates glucagon-induced delayed gastric emptying in dogs

Am J Surg. 2008 Jan;195(1):122-9. doi: 10.1016/j.amjsurg.2007.04.007.

Abstract

Background: The aim of the current study was to investigate the efficacy of 2-channel gastric electrical stimulation (GES) with a custom-made implantable pacemaker on delayed gastric emptying and gastric dysrhythmia induced by glucagon in dogs.

Methods: Six dogs were studied in 4 randomized session (saline, glucagon, glucagon with single-channel or 2-channel GES). GES was applied via the first pair of electrodes for single-channel GES or the first and third pairs of electrodes for 2-channel GES. Gastric emptying was assessed for 90 minutes and gastric slow waves were recorded at the same time.

Results: Both single-channel and 2-channel GES improved gastric dysrhythmia (P < .05 vs glucagon session). Two-channel GES but not single-channel GES improved glucagon-induced delayed gastric emptying at 30 minutes, 45 minutes, 60 minutes, 75 minutes, and 90 minutes.

Conclusion: Two-channel GES with a novel implantable pacemaker is more efficient and effective than single-channel GES in improving delayed gastric emptying induced by glucagon. This implantable multipoint pacemaker may provide a new option for treatment of gastric motility disorders.

Publication types

  • Research Support, N.I.H., Extramural

MeSH terms

  • Animals
  • Blood Glucose / analysis
  • Dogs
  • Electric Stimulation
  • Electric Stimulation Therapy*
  • Electrodes, Implanted
  • Female
  • Gastric Emptying*
  • Gastroparesis / chemically induced
  • Gastroparesis / therapy*
  • Glucagon / adverse effects
  • Treatment Outcome

Substances

  • Blood Glucose
  • Glucagon