Incidental testicular irradiation from prostate IMRT: it all adds up

Int J Radiat Oncol Biol Phys. 2010 Jun 1;77(2):484-9. doi: 10.1016/j.ijrobp.2009.04.083. Epub 2009 Sep 3.

Abstract

Purpose: To identify the technical aspects of image-guided intensity-modulated radiation therapy (IMRT) for localized prostate cancer that could result in a clinically meaningful incidental dose to the testes.

Methods and materials: We examined three sources that contribute incidental dose to the testes, namely, from internal photon scattering from IMRT small field and large pelvic nodal fields with 6 or 15 MV, from neutrons when >10-MV photons are used, and from daily image-guided fiducial-based portal imaging. Using clinical data from 10 patients who received IMRT for prostate cancer, and thermo-luminescent dosimeter measurements in phantom, we estimated the dose to the testes from each of these sources.

Results: A mean testicular dose of 172 and 220 cGy results from internal photon scatter for pelvic nodal fields and 68 and 93 cGy for prostate-only fields, for 6- and 15-MV energies, respectively. For 15-MV photon energies, the mean testicular dose from neutrons is 60 cGy for pelvic fields and 31 cGy for prostate-only fields. From daily portal MV image guidance, the testes-in-field mean dose is 350 cGy, whereas the testes-out-of-field scatter dose is 16 cGy. Dosimetric comparisons between IMRT using 6-MV and 15-MV photon energies are not significantly different. Worst-case scenarios can potentially deliver cumulative incidental mean testicular doses of 630 cGy, whereas best-case scenarios can deliver only 84 cGy.

Conclusions: Incidental dose to the testes from prostate IMRT can be minimized by opting to restrict the use of elective pelvic nodal fields, by choosing photon energies <10 MV, and by using the smallest port sizes necessary for daily image guidance.

MeSH terms

  • Humans
  • Leydig Cells / radiation effects
  • Male
  • Neutrons / therapeutic use
  • Photons
  • Prostatic Neoplasms / radiotherapy*
  • Radiometry / methods
  • Radiotherapy Dosage
  • Radiotherapy Planning, Computer-Assisted / methods*
  • Radiotherapy, Intensity-Modulated / methods*
  • Scattering, Radiation*
  • Testis / radiation effects*
  • Thermoluminescent Dosimetry / methods