Improved dose-calculation accuracy in proton treatment planning using a simplified Monte Carlo method verified with three-dimensional measurements in an anthropomorphic phantom

Phys Med Biol. 2010 Jun 21;55(12):3545-56. doi: 10.1088/0031-9155/55/12/018. Epub 2010 May 28.

Abstract

Treatment planning for proton tumor therapy requires a fast and accurate dose-calculation method. We have implemented a simplified Monte Carlo (SMC) method in the treatment planning system of the National Cancer Center Hospital East for the double-scattering beam delivery scheme. The SMC method takes into account the scattering effect in materials more accurately than the pencil beam algorithm by tracking individual proton paths. We confirmed that the SMC method reproduced measured dose distributions in a heterogeneous slab phantom better than the pencil beam method. When applied to a complex anthropomorphic phantom, the SMC method reproduced the measured dose distribution well, satisfying an accuracy tolerance of 3 mm and 3% in the gamma index analysis. The SMC method required approximately 30 min to complete the calculation over a target volume of 500 cc, much less than the time required for the full Monte Carlo calculation. The SMC method is a candidate for a practical calculation technique with sufficient accuracy for clinical application.

MeSH terms

  • Humans
  • Monte Carlo Method*
  • Phantoms, Imaging*
  • Proton Therapy*
  • Radiometry / instrumentation*
  • Radiotherapy Planning, Computer-Assisted / instrumentation*

Substances

  • Protons