Residual translational and rotational errors after kV X-ray image-guided correction of prostate location using implanted fiducials

Strahlenther Onkol. 2010 Oct;186(10):544-50. doi: 10.1007/s00066-010-2030-8. Epub 2010 Sep 30.

Abstract

Purpose: To evaluate the residual errors and required safety margins after stereoscopic kilovoltage (kV) X-ray target localization of the prostate in image-guided radiotherapy (IGRT) using internal fiducials.

Patients and methods: Radiopaque fiducial markers (FMs) have been inserted into the prostate in a cohort of 33 patients. The ExacTrac/Novalis Body™ X-ray 6d image acquisition system (BrainLAB AG, Feldkirchen, Germany) was used. Corrections were performed in left-right (LR), anterior-posterior (AP), and superior-inferior (SI) direction. Rotational errors around LR (x-axis), AP (y) and SI (z) have been recorded for the first series of nine patients, and since 2007 for the subsequent 24 patients in addition corrected in each fraction by using the Robotic Tilt Module™ and Varian Exact Couch™. After positioning, a second set of X-ray images was acquired for verification purposes. Residual errors were registered and again corrected.

Results: Standard deviations (SD) of residual translational random errors in LR, AP, and SI coordinates were 1.3, 1.7, and 2.2 mm. Residual random rotation errors were found for lateral (around x, tilt), vertical (around y, table), and longitudinal (around z, roll) and of 3.2°, 1.8°, and 1.5°. Planning target volume (PTV)-clinical target volume (CTV) margins were calculated in LR, AP, and SI direction to 2.3, 3.0, and 3.7 mm. After a second repositioning, the margins could be reduced to 1.8, 2.1, and 1.8 mm.

Conclusion: On the basis of the residual setup error measurements, the margin required after one to two online X-ray corrections for the patients enrolled in this study would be at minimum 2 mm. The contribution of intrafractional motion to residual random errors has to be evaluated.

MeSH terms

  • Algorithms
  • Cohort Studies
  • Contrast Media*
  • Fiducial Markers / statistics & numerical data
  • Humans
  • Male
  • Prostatic Neoplasms / diagnostic imaging*
  • Prostatic Neoplasms / radiotherapy
  • Prostheses and Implants
  • Radiotherapy Planning, Computer-Assisted / adverse effects
  • Radiotherapy Planning, Computer-Assisted / methods
  • Tomography, X-Ray Computed / methods
  • Ultrasonography
  • X-Rays